Tubu Furnace
Two Zone Split Tube Furnace 1200°C
Two-zone split tube furnaces are typically used for applications that require precise temperature control or for applications where the use of different atmospheres is necessary. Some common applications for two-zone split tube furnaces include:
Two-zone split tube furnaces are available in a variety of sizes and temperature ranges. The size of the furnace will determine how many samples can be heated at once, while the temperature range will determine the types of materials that can be heated in the furnace.
Application
1200°C Two Zone Split Tube Furnace Mainly Provides High Temperature Heat Treatment Environment for Industrial, Research Institutes, Factories and Other Industrial Laboratories, And Is Applied to New Materials Such As Metal Materials, Ceramic Materials, Nano Materials, And Semiconductor Materials.
Technical Specifications
Type
Two Zone Split Type
Max Temperature Zone 1
1200℃
Max Temperature Zone 2
1200℃
Continue Operating Temp. Zone 1
1150℃
Continue Operating Temp. Zone 2
1150℃
Heating Element
Kanthal “A-1” Wire in Coil Form
Insulation
➢ First layer: 1430℃ ceramic fiber board& Zirconia Fiber Blanket.
➢ Second layer: 1260℃ ceramic fiber board & Zirconia Fiber Blanket.
➢ Third Layer: High Quality Ceramic Fiber Zirconia Grade Thermal Insulation Material To Ensure A Good Uniformity.
Temperature Accuracy
±1℃
Accuracy
Thermocouple
K type (Pt-Rh to Pt-Rh) with 99.7% purity Alumina tube for Each Zone
Temperature Controller
➢ PID automatic control via SSR power control with phase angle fired, current limiting resistor.
➢ 1 programmable x 16 segments for precise control of heating rate, cooling rate and dwell time.
➢ Built in PID Auto-Tune function with overheating & broken thermocouple broken protection for this Split tube Furnace 1200 Type.
➢ Over temperature protection and alarm allows for operation without attendant
Electronic Component
Autonics / Schneider / L&T Electronics Brands
Heating Rate
≤ 15℃/Min (Suggest 10℃/Min for Longer Life Using of Furnace)
Safety Protection
Overheat And Thermocouple-Break Alarm
Certification
NSIC REGD, MSME REGD, ISO 9001 – 2015/14001-2015/13485-2016/FDA/GMP/CE CERTIFIED
Furnace Shell
High Quality GI Sheet With PP Powder Coated
Power Supply
220 V, Single Phase AC SUPPLY
Warranty
➢ One-year limited warranty for this Split Tube Furnace 1200 type with lifetime support. (Consumable parts such as heating elements and Sensor, Tube, crucibles are not covered by the warranty, please order the replacement at related products)
➢ ATTENTION: Any damages caused by the use of corrosive and acidic gases are not under the coverage of One Year Limited Warranty.
Using Attentions
➢ The cooling rate of this Split Tube Furnace1200 type shall also not exceed 10℃/min.
➢ Toxic or explosive gases are not recommended for use with this Two Zone Split Tube Furnace without necessary safety controls and supervision.
➢ Small cracks may appear on the surface of the refractory ceramics over extended use. This is a normal occurrence
Send Message